| C.A 8220 firmware 1.4 (2011-09-26) |
+------------------------------------+

The embedded software of the C.A 8220 can be updated by the user
with the optical link.

Warning: updating of the embedded software causes all data to be 
deleted (configuration, photographs).
Back up any data to be kept on a PC using the dedicated software 
before updating the embedded software.

Updates to the embedded software depend on its compatibility with 
the hardware version of the instrument.

Displaying information
The serial number, the software version and the hardware version of 
the C.A 8220 can be displayed on the screen. 
Proceed as follows:
1. Switch the instrument to OFF. Turn the rotary switch to OFF.
2. Switch the instrument back ON. Turn the rotary switch to any position. 
   When the first screen is displayed, press the white key and keep it 
   pressed down until the first information page is displayed.
3. You can navigate among the information pages by using the Right / Left arrow keys.  

+-------------------+
| Upgrade procedure |
+-------------------+
1) Download the compress file
This ZIP archive contains the following files :
	- hex55.exe
	- CA8220.out
	- PloaderCA8220.exe
	- ReadMe-First.txt
	- LisezMoi-dabord.txt

2) All those files must be extracted in the same directory
3) Run "PloaderCA8220.exe"
4) Follow the instructions...

Note : Be sure to have your instrument correctly linked to your PC.

+------------------+
| Revision history |
+------------------+

Firmware 1.4
------------
- (change) Changing the detection mechanism OUT OF RANGE when no sensor is detected.


Firmware 1.3
------------
- (additional feature) Adding the devices compatibility with the E3N current sensor, coupled with the appropriate adapter.
- (correction) The counting of energy is not arrested in case of "out of range".

Note : this firmware is compatible with hardware versions 0.3 and 1.3.

Firmware 1.2
------------
- (correction) The sign of the reactive values is now correctly handled in 3PHI mode with harmonics.
- (correction) The Rogowski-current-channel gain is set to 1 in Inrush mode (formerly set to 0).
- (correction) Energy counters are now all positive (absolute values).

Note : this firmware is compatible with hardware versions 0.3 and 1.3.

Firmware 1.1
------------
- (additional feature) Energy counting with 8 independent counters (10 pages have been added in "power" mode).
- (additional feature) The DC value of the current is now displayed (in calibration mode) in "current harmonic" mode.
- (correction) An error test for "VAR-with-harmonics" calculation has been added.
- (correction) The "Tangent-Phi" error conditions have been corrected.
- (correction) The "Wave.AFrms" value is now never forced to zero (in order to avoid a frequency calculation instability).
- (change) Frequency error calculation propagation is now deeper.

Note : this firmware is compatible with hardware versions 0.3 and 1.3.

Firmware 1.0
------------
Initial revision.

Note : this firmware is compatible with hardware versions 0.3 and 1.3.